Applicable Syringes | 5ml,10ml,20ml,30ml,50ml,60ml of any standard |
Accuracy | +/ - 2% with correct calibration |
Rate Mode 50ml(60ml) | 0.1ml/h - 2000.0ml/h (0.1ml/h or 1ml/h step) |
Rate Mode 30ml | 0.1ml/h - 1200.0ml/h (0.1ml/h or 1ml/h step) |
Rate Mode 20ml | 0.1ml/h - 800.0ml/h (0.1ml/h or 1ml/h step) |
Rate Mode 10ml | 0.1ml/h - 400.0ml/h (0.1ml/h or 1ml/h step) |
Rate Mode 5ml | 0.1~200ml/h (0.1ml/h or 1ml/h step) |
Rate Mode (Programmable) | 0.01ml/h~9999.9ml/h |
Time Mode SOL. VOL | 0.1 - 9999.9ml (0.1ml step) |
Time Mode TIME | 00:01 - 99:59min (1min step) |
Dose Mode Volume | 0.1 - 999.9ml (0.1ml step) |
Dose Mode DRUG MASS | 0.1 - 999.9mg ( 0.1mg step) |
Dose Mode DOSE | 0.001 - 9999ug/kg/min (0.001ug/kg/min step) |
Dose Mode WEIGHT | 0.1 - 300kg (0.1kg the least step) |
Dose unit | ng/kg/min, ug/kg/min, mg/kg/min, g/kg/min, ug/kg/h, mg/kg/h, g/kg/h, ug/min, mg/min, g/min, ug/h, mg/h, g/h, U/min, U/h, IU/h, U/kg/h, ml/h |
Total Volume Range | 0.1 - 9999.9ml (0.1 ml step) |
Drug library Mode | 10 categories of drugs, 17kinds of dose units. Makes drug dispensing more convenient and safer. |
Sequential Mode | Available |
Purge Rate 50ml(60ml) | 1500.0ml/h |
Purge Rate 30ml | 1200.0ml/h |
Purge Rate 20ml | 800.0ml/h |
Purge Rate 10ml | 400.0ml/h |
Purge Rate 5ml | 200ml/h |
Purge Rate (Programmable and adjustable) | - |
KVO Rate | 0.1~5.0 ml/h, Programmable and adjustable |
DPS Dynamic Pressure | Three levels setting LOW (60(± 15)kPa), MEDIUM (90(± 15)kPa) or HIGH (120(± 15) kPa) |
Alarms | Syringe Holder Loose, Syringe Plunger Disengagement, Syringe Flange Insertion Error, Syringe Not Calibrated, Occlusion, Syringe Empty, Rate Abnormal, Battery Empty, No Battery, Infusion Complete, Syringe Near Empty, No Action, Battery Low |
Alarm volume | Three levels of alarm volume LOW, MEDIUM or HIGH |
Bolus Rate 5ml | 0.1~200ml/h |
Bolus Rate 10ml | 0.1 - 400ml/h |
Bolus Rate 20ml | 0.1 - 800ml/h |
Bolus Rate 30ml | 0.1 - 1200ml/h |
Bolus Rate 50ml(60ml) | 0.1 - 2000ml/h |
Bolus Rate (Programmable and adjustable, Automatic/Manual bolus available) | - |
Bolus Volume Range | 1.0~50.0ml, (0.1ml step) |
Interface | RS - 232 |
Brightness | 1~10 levels can be selected |
History Records | 2000 events stored can be checked and downloaded |
Power Source | AC power supply: AC 100 to 240V(+/ - 10%), 50/60Hz; DC Power supply: 12~15V, power>15W; Built - in
Lithium
Battery: Rechargeable, 7.2V, 2600mAh. 8 hours of continuous operation |
Protection Class | Class I, Type CF, IP44 |
Dimension/N.W | 350*150*220mm, 4kg |
Packing size/G.W | 400mm*200mm*2500mm, 4.5KG |