Characteristics
Door Safety Interlock
Steam injection is automatically disabled when the door is unsealed
Visual indicator shows dry cycle operation availability
Pressure-Activated Door Lock
Door automatically locks when internal pressure exceeds 0.027MPa
Remains locked until safe pressure levels are restored
Cycle Initiation Requirements
Audible alarm activates if sterilization time isn't programmed before startup
Mid-Cycle Door Protection
Automatic shutdown of jacket valve and opening of chamber valve if door is opened during operation
Water Level Monitoring System
Activates warning light when water is insufficient
Automatically cuts heating/sterilization power
Triggers audible alarm and steam release mechanism
Electrical Safety Protections
Electrical leakage incidents
Current overload conditions
Automatic power cutoff during
Pressure Fail-Safe Mechanism
Safety valve automatically activates at 0.165MPa if temperature control fails
Construction Specifications
Complete stainless steel construction for durability and hygiene
Technical Data
Model technical data | WS-280YDC | WS-200YDC | WS-150YDC |
Sterilizing chamber volume | 280L(φ600×1000mm) | 200L(φ515×1000mm) | 150L(φ440×1000mm) |
working pressure | 0.22 MPa | ||
working temperature | 134℃ | ||
Adjustment of temperature | 40℃ - 134℃ | ||
Time of sterilization | 0-60 min | ||
Time of drying | 0-60 min | ||
Heat average | ≤± 2℃ | ||
Power | 12KW/380V 50Hz | 9KW/380V 50Hz | 9KW/380V 50Hz |
Dimension | 1400×770×1780 | 1400×670×1650 | 1400×600×1300 |
Package | 1680×920×2100 | 1560×820×1850 | 1550×750×1850 |
GW/NW | 465/365 KG | 350/260 KG | 320/240 KG |