H1200-1 Anestezi Cihazı H1200-1 - Anaesthesia Machine | HQS Medical
Anestezi Cihazı

H1200-1 Anestezi Cihazı

Model: H1200-1

Giriş — H1200-1 Anestezi Cihazı

HQS Vetcare delivers a high-precision veterinary anesthesia equipment system suitable for animal hospitals, pet clinics, and animal laboratories. This anesthesia machine meets the needs of general anesthesia and medical research for various animals, including mice, dogs, cats, rabbits, monkeys, pigs, and sheep.

Temel Özellikler — H1200-1

  • Application 0.5 - 100kg animal
  • Ventilation mode Closed, half - closed, half - open
  • Tidal volume 10 - 2000ml
  • Vaporizer Choose one from Isoflurane, Sevoflurane, Halothane
  • Oxygen source pressure 0.25~0.65Mpa
  • Airway pressure scope -30~60CMH2O
  • Flow Control 0 - 4L/min
  • Oxygen Flush 35L/min~75L/min
  • Pop adjust scope Directs the waste anesthetic gas from the machine to a scavenging system. In the completely open position, the pop - off valve will release pressure at 2 cm H2O, while maintain a constant passive volume in the breathing bags.
  • Inspiration impedance ≦ 0.6Kpa
  • Expiration impedance ≦ 0.6Kpa
  • Absorption tank capacity 700ml
  • Monitoring Airway pressure, Gas source pressure monitoring
  • I:E As doctor need
  • BPM As doctor need
  • Optional V-60V ventilator – Pneumatically driven and electrically controlled
  • Display: LCD digital display
  • Animal Bellow : Big animal : 300 - 1500ML ; Small animal : 0 - 300ML
  • Monitoring : Tidal volume, breathing rate, I:E ratio, p eak airway pressure
  • TV: 20 - 1500ml
  • Breathing rate: 4~60bpm
  • I:E ratio: Adjustable 3:1,2:1,1.5:1,1:1,1:1.5,1:2,1:3
  • Upper airway pressure:2~6kpa
  • Lower airway pressure:0.6~2kpa
  • Build in battery : work time more than 2 hours in case of power failure
  • Alarm: Battery low, Power failure, Airway pressure high/low
  • Alarm Silence : Manuel/Auto Switch
  • PIP scope : 5 - 60 cmH2O
  • Optional V-80V ventilator – Display: 9-inch TFT touch screen
  • Drive gas: Gas type: O2 or Air
  • Inlet pressure: 280 - 600 kPa
  • Max flow : ≤ 120 L/min
  • Operating mode : VCV, PCV, SPONT, Manual, Demo
  • Waveform: Pressure, Flow - rate, Volume Loop: PV loop, PF loop, FV loop
  • Safety pressure: System pressure does not exceed 6 KPa
  • Battery type: Li - ion, work time more than 2 hours
  • Tidal volume: 20 - 1500ml; Increment: 1 ml
  • Respiratory rate: 1 ~ 100 bpm
  • Inspiratory time: 0.1 ~ 10s
  • Respiratory ratio: 9.9:1 to 1:9.9
  • Percentage of inspiratory pause: 0 to 50%
  • PEEP: OFF, 3 ~ 20 cmH2O
  • Pressure Support: 0 ~ 60 cmH2O
  • Pressure Control: 5 ~ 60 cmH2O
  • Flow trigger : 0.5 ~ 20L/min
  • Pressure Trigger Range: - 1 ~ - 20 cmH2O
  • PSV Insp Termination Level : 25%
Fiyat listesini e-postayla alın

E-posta adresinizi bırakın; bu modelin teklifini ve teknik veri sayfasını genellikle bir iş günü içinde gönderelim.

Share:

Model-specific documentation

Documents for H1200-1

Check the covered model, language and document status before use. The manual supplied with the actual unit remains the controlling reference.

Product datasheet · English

H1200-1 H1200-1 Anesthesia Machine product datasheet

Public PDF

HQS datasheet for the H1200-1: product overview, features and the technical specifications published for this model. Confirm the ordered configuration in your quotation.

  • Product overview
  • Features
  • Technical specifications
Covered model
H1200-1
Length
3 pages
File
PDF · 314 KB
Web copy prepared
Need a different language or configuration-specific document?

H1200-1 Anestezi Cihazı Teknik Özellikler

PARAMETER
SPECIFICATION
Application
0.5-100 kg animal
Ventilation mode
Closed, half-closed, half-open
Tidal volume
10-2000 ml
Vaporizer
Choose one from Isoflurane, Sevoflurane, Halothane
Oxygen source pressure
0.25-0.65 MPa
Airway pressure scope
-30 to 60 cmH2O
Flow control
0-4 L/min
Oxygen flush
35-75 L/min
Pop-off adjust scope
Directs waste anesthetic gas to a scavenging system; fully open, the pop-off valve releases pressure at 2 cmH2O while keeping a constant passive volume in the breathing bag
Inspiration impedance
≤0.6 kPa
Expiration impedance
≤0.6 kPa
Absorption tank capacity
700 ml
Monitoring
Airway pressure, gas source pressure
I:E and BPM
Set by the veterinarian

Konfigürasyon

  • Main unit, one anesthesia vaporizer, one set veterinary breathing circuit, breathing circuit, gas source hose, exhaust gas absorption system, corrugated connect pipeline, simulated lung.
  • Optional: V-60V or V-80V anesthesia ventilator

Seçenekler ve aksesuarlar

  • Anesthesia ventilator (V - 60V, V - 80V)
  • trolley
  • non - breathing circuit
  • tracheal intubation
  • anesthesia laryngoscope
  • soda lime
  • active carbon

H1200-1 için garanti ve satış sonrası hizmet

HQS bu model için garanti, yedek parça ve onarımı nasıl yürütür.

Garanti süresi
Ana ünite garantisi modele ve siparişe göre değişir. Geçerli süre teklifinizde ve sipariş onayınızda belirtilir; lütfen satın almadan önce teyit edin.
Kapsam
Garanti süresi içinde üretim veya kalite kusurundan kaynaklanan bir arıza ücretsiz onarılır. Sarf malzemeleri ve aksesuarlar, hatalı kullanımdan doğan hasarlar ve seri numarası etiketi değiştirilmiş ya da çıkarılmış cihazlar ana ünite garantisi kapsamında değildir.
Yedek parça
Yedek parça ve teknik destek, HQS’nin ürünü sevk ettiği tarihten itibaren en az sekiz yıl sağlanır.
Yanıt süresi
Servis taleplerine bir iş günü içinde ilk yanıt verilir. Lütfen seri numarasını, modeli, ana ünite sürümünü ve sorunun açıklamasını gönderin.
Garanti içi onarım
Onaylanmış bir garanti arızası yedek parça gerektiriyorsa HQS parçayı kendi masrafıyla gönderir; arızalı parça HQS’ye iade edilir.
Garanti sonrası onarım
HQS önce maliyet bazlı bir teklif ve proforma fatura gönderir. Onarılan parçalar iade edilmeden önce test edilir. E-posta veya telefonla teknik destek ücretsizdir.

Garanti ve hizmet koşullarının tamamı →