H1200-3 Anesthesia Machine H1200-3 - Anaesthesia Machine | HQS Medical
Anaesthesia Machine

H1200-3 Anesthesia Machine

Model: H1200-3

Introduction — H1200-3 Anesthesia Machine

HQS Vetcare delivers high-precision veterinary anesthesia equipment with a ventilation management system, suitable for animal hospitals, pet clinics, and animal laboratories. This anesthesia machine meets the needs of general anesthesia and medical research on a wide range of animals.

Key Features — H1200-3

  • Ventilation manage system
  • Pop off valve directs the waste anesthetic gas from the machine to a scavenging system.
  • In the completely open position, the pop - off valve will release pressure at 2 cm H2O, while maintain a constant passive volume in the breathing bags.
Get the price list by email

Leave your email and we'll send the quotation and datasheet for this model, usually within one working day.

Share:

Model-specific documentation

Documents for H1200-3

Check the covered model, language and document status before use. The manual supplied with the actual unit remains the controlling reference.

Product datasheet · English

H1200-3 H1200-3 Anesthesia Machine product datasheet

Public PDF

HQS datasheet for the H1200-3: product overview, features and the technical specifications published for this model. Confirm the ordered configuration in your quotation.

  • Product overview
  • Features
  • Technical specifications
Covered model
H1200-3
Length
2 pages
File
PDF · 250 KB
Web copy prepared
Need a different language or configuration-specific document?

H1200-3 Anesthesia Machine Specifications

PARAMETER
SPECIFICATION
Application
0.5 - 100kg animal
Respiration mode
IPPV, Manuel
Ventilation mode
Open , c losed, half - closed, half - open
Tidal volume
20 - 1600ml (machine control); 5 - 1600ml (manual)
Vaporizer
Choose one from Isoflurane, Sevoflurane, Halothane
Oxygen source pressure
0.25~0.65Mpa
Gas source pressure monitoring scope
0 - 1Mpa
Airway pressure scope
- 30~60CMH2O
Flow Control
0 - 5L/min
Oxygen Flush
25L/min~75L/min
Inspiration impedance
≦ 0.6Kpa
Expiration impedance
≦ 0.6Kpa
Absorption tank capacity
700ml
Monitoring
Tidal volume, Airway pressure, Gas source pressure monitoring, breathing rate I:E ratio I : E Adjustable: 3:1,2:1,1.5:1,1:1,1:1.5,1:2,1:3
BPM
2~150bpm
Bellow
Big animal t:50 - 1600ml, small animal:0 - 300ml

Configuration

  • Main unit
  • Anesthesia ventilator
  • One set veterinary breathing circuit
  • Breathing circuit
  • Gas source hose
  • Exhaust gas absorption system
  • Corrugated connect pipeline
  • Simulated lung
  • One anesthesia vaporizer (isoflurane, sevoflurane or halothane)

Options & accessories

  • Trolley
  • Non-breathing circuit
  • Tracheal intubation
  • Anesthesia laryngoscope
  • Soda lime
  • Active carbon

Buyer decision support

H1200-3 anaesthesia machine with ventilator procurement decision notes

The H1200-3 is the HQS veterinary anaesthesia machine with an IPPV ventilator in the standard configuration, for animals from 0.5 to 100 kg. It supports open, closed, half-closed and half-open circuits, works from oxygen at 0.25–0.65 MPa with a 0–5 L/min flow control and 25–75 L/min oxygen flush, and takes one isoflurane, sevoflurane or halothane vaporizer. The ventilator offers IPPV and manual modes, tidal volume 20–1600 ml in machine control or 5–1600 ml manually, 2–150 bpm, and I:E steps of 3:1, 2:1, 1.5:1, 1:1, 1:1.5, 1:2 and 1:3, with big-animal (50–1600 ml) and small-animal (0–300 ml) bellows. It monitors tidal volume, airway pressure (−30 to 60 cmH2O), gas source pressure (0–1 MPa), breathing rate and I:E. The absorber holds 700 ml and inspiratory and expiratory resistance are each ≤0.6 kPa. Within the H1200 series it sits between the H1200-5 (4–60 bpm, up to 1500 ml) and the touch-screen H1200-2 (PCV, VCV, PEEP); choose it when you need a wide breathing-rate range for small patients without the touch-screen ventilator.

Decision factorPublished for this modelConfirm before order
Patients and circuits0.5–100 kg animals; open, closed, half-closed and half-open; oxygen 0.25–0.65 MPa; flow control 0–5 L/min; oxygen flush 25–75 L/minTell us your patient range and oxygen source, so we can confirm circuits and hoses.
VentilatorIPPV and manual; tidal volume 20–1600 ml (machine control), 5–1600 ml (manual); 2–150 bpm; I:E 3:1, 2:1, 1.5:1, 1:1, 1:1.5, 1:2, 1:3; bellows 50–1600 ml (big animal) and 0–300 ml (small animal)Ask which bellows are supplied as standard for your patient sizes.
MonitoringTidal volume, airway pressure (−30 to 60 cmH2O), gas source pressure (0–1 MPa), breathing rate, I:E ratioAsk for the alarm list in the operation manual if you need specific alarms.
Vaporizer and absorberOne vaporizer: isoflurane, sevoflurane or halothane; absorber 700 ml; inspiration and expiration resistance ≤0.6 kPa; pop-off valve releases at 2 cmH2O fully openChoose the anaesthetic agent and ask about the vaporizer mounting type.
ConfigurationStandard: main unit, anaesthesia ventilator, one vaporizer, breathing circuit, gas source hose, exhaust gas absorption system, corrugated pipeline, test lung. Optional: trolley, non-rebreathing circuit, endotracheal tubes, laryngoscope, soda lime, activated carbonList the optional items you want in the quotation.

Prepared by HQS Medical Product Team · Documentation reviewed:

Source: H1200-3 Anesthesia Machine model brochure. Verify the final configuration, documents and destination-market requirements in the quotation and supplied manual.

Warranty and after-sales service for H1200-3

How HQS handles warranty, spare parts and repairs for this model.

Warranty period
The main-unit warranty varies by model and order. The period that applies is stated in your quotation and order confirmation; please confirm it before purchase.
What is covered
A malfunction caused by a manufacturing or quality defect within the warranty period is repaired free of charge. Consumables and accessories, damage from misuse, and units with an altered or removed serial label are not covered by the main-unit warranty.
Spare parts
Spare parts and technical support are available for at least eight years from the date HQS ships the product.
Response time
Service requests get a first reply within one business day. Please send the serial number, model, main-unit version and a description of the problem.
Repairs in warranty
If a confirmed warranty fault needs a replacement part, HQS ships the part at its own expense, and the faulty part is returned to HQS.
Repairs after warranty
HQS sends a cost-based quotation and proforma invoice first. Repaired parts are tested before they are returned. Technical help by email or phone is free.

Full warranty and service terms →